AL2021 hydrogen peroxide analyzer, front view
Continuous H₂O₂ Analyzer

AL2021

Continuous hydrogen peroxide measurement in air and water down to 100 ppt — optimized versions for pharma and semiconductor manufacturing (UPW).

The AL2021 continuously measures hydrogen peroxide (H₂O₂) using a highly selective enzymatic reaction based on peroxidase. Its exceptional selectivity and sensitivity make it an ideal choice for monitoring hydrogen peroxide at concentrations down to the low ppb range. In addition, the method allows straightforward calibration and provides fully traceable measurement results.

These advantages enable a wide range of applications, including atmospheric research, the pharmaceutical and beverage industries, as well as water quality monitoring. The AL2021SCL is particularly well suited for liquid samples and is a fully automatic H₂O₂ analyzer capable of continuous measurements in the low µg/L range. This makes it an important tool for semiconductor manufacturers who need to ensure the quality of ultrapure water (UPW).

Calibratable, Traceable, Reliable
The instrument can be easily calibrated by the user with a liquid standard. All steps and calculations are traceable.
Sub-PPB in Both Phases
<100 ppt in gas and <100 ng/l in liquid, linear across four orders of magnitude.
Documented Cross-Interferences
Only two significant ones: O₃ < 1:3500 and NO < 1:8000. We publish them rather than hide them.
Runs Unattended
Automatic zeroing through the internal zero trap, optional automatic calibration from an internal permeation source.
Read more: Technical Details →

Technical Specifications

All specifications at standard conditions (23 °C, 1013 hPa, dry gas).

Measurement
AL2021 — Hydrogen Peroxide (H₂O₂)
Detection PrincipleEnzymatic UV fluorimetry (peroxidase / p-hydroxyphenylacetic acid), continuous
Measurement Range0.1 ppb – 3000 ppb (gas) · 100 ng/l – 3 mg/l (liquid)
Response Time (T90)~90 s (T90), delay ~300 s
Accuracy2% full scale
Sample Flow2 L/min air · ~1.0 mL/min liquid
Channels / Chambers2 channels (A = all peroxides, B = without H₂O₂) · multiple gas sources can be connected via valve cluster
Cross-Interferences
O₃ < 1:3500 · NO < 1:8000. Other peroxides are detected as a relative signal only. No detectable interference: acetone, methylamine, dimethylamine, n-butane, cis-/trans-2-butene, SO₂, PAN, NO₂, glyoxal, isobutane, isobutylene, 1-butene, HCHO, benzene, toluene, methanol; ions such as iodide, chloride, nitrate, bromide, phosphate, benzoate.
Instrument
General
Dimensions (W × D × H)45 × 15 × 56 cm (19″)
Weight15 kg
Power Supply110 / 230 VAC, 50/60 Hz · 24 VDC on request
Power Consumption110 W
Warm-up Time15 minutes
Analog Outputoptional
Digital InterfaceRS232 · TCP/IP (RS232 ↔ USB adapter included)

Interested in the AL2021?