AL5005 fast carbon monoxide analyzer with tablet, front view
Ultra-Fast CO Analyzer

AL5005

Ultra-fast carbon monoxide measurement below 1 ppb with 0.1 s real time response — for airborne campaigns and long-term background monitoring.

The AL5005 has a wide range of applications in environmental and climate research, where accurate real-time CO measurements provide valuable insights into pollution sources, atmospheric processes, and climate dynamics. With a time resolution of 100 ms, the instrument is particularly well suited for airborne measurements. During research flights, high temporal resolution translates directly into high spatial resolution, allowing rapid changes in CO concentrations and small-scale atmospheric structures to be captured precisely.

The AL5005 is also well suited for remote research environments, such as Antarctica, where its robustness and stand-alone operation ensure reliable performance under demanding conditions. These capabilities make it a versatile instrument for field campaigns and long-term atmospheric monitoring at remote sites.

100 ms in Real Time
No other CO analyzer measures in real time at 100 ms. On an aircraft, time resolution becomes spatial resolution.
No Cross-Sensitivities
There are no known interferences except water vapour, which is removed with a Nafion dryer.
A Reference Instrument
VUV fluorimetry is regarded as the reference method for CO. NOAA, GAW and the Chinese Academy of Sciences use the AL5005 as a reference.
Five Orders of Magnitude, One Calibration
Linear from 1 ppb to 100 ppm. Empa Dübendorf verified the linearity of the measuring principle on the AL5001/5002.
Built for Places You Cannot Visit
Rugged, stand-alone, 110/230 V AC or 24 V DC on request, USB logging, internal zero source. Sample pressure 200–1200 mbar, i.e. up to about 12000 m.
Calibrates Itself in Minutes
Fully automatic, schedulable, on a small amount of calibration gas.
Read more: Technical Details →

Technical Specifications

All specifications at standard conditions (23 °C, 1013 hPa, dry gas).

Measurement
AL5005 — Carbon Monoxide (CO)
Detection PrincipleVUV resonance fluorescence, continuous
Measurement Range1 ppb – 100 ppm
Extended RangeExtendable up to 1:100 with AL GD2 gas diluter (linearity verified with AL5005)
Time Resolution0.1 s real time (stronger scroll pump required at high altitude)
Accuracy2% full scale
Cross-Interferences
No known interferences except water vapour, which is removed with a Nafion dryer.
Instrument
General
Dimensions (W × D × H)45 × 23 × 56 cm (19″)
Weight22 kg
Power Supply110/230 VAC · 24 VDC on request · adapted to the aircraft for airborne use
Digital InterfaceRS232 · TCP/IP (RS232 ↔ USB adapter included)
Funding

Interested in the AL5005?

Speak directly to the engineer responsible for this instrument — no sales detours.