AL2021SC hydrogen peroxide analyzer, front view
Continuous H₂O₂ Analyzer for Pharma

AL2021SC

Continuous hydrogen peroxide measurement in gas and liquid samples down to 100 ppt — the single-channel version for isolators and RABS in aseptic filling.

The AL2021SC continuously measures hydrogen peroxide (H₂O₂) using a highly selective enzymatic reaction based on peroxidase. It is the single-channel version of the AL2021, developed for the filling of injectables such as vaccines and enzymes in isolators and RABS.

Isolators are decontaminated with vaporized H₂O₂. This process is generally monitored with electrochemical sensors that are sensitive down to about 1 ppm, which is far too high for many vaccines or enzymes. The AL2021SC closes this gap with near real-time data on the H₂O₂ concentration. It measures at the outlet gas stream of the isolator or RABS, is not connected to its interior and therefore does not need to be sterile itself.

Starts When the Electrochemical Sensor Reaches Its Limits
Standard H₂O₂ sensors of isolators are designed to measure concentrations above 1 ppm. For sensitive pharmaceutical products, this concentration is far too high. The AL2021SC enables the user to monitor concentrations in a range of 0–3 ppm. This makes it the perfect add-on for the isolator.
Gas and Liquid in One Instrument
Unlike other instruments, the AL2021SC handles both gas and liquid samples. This makes it easy to measure the uptake of H₂O₂ from the isolator air into the injectable during filling.
Absolute Concentration, Not a Relative Signal
The AL2021SC measures peroxides by an enzymatic reaction with p-hydroxyphenylacetic acid and reports an absolute H₂O₂ concentration, traceable to a liquid standard. Detection limit below 100 ppt.
Runs Unattended
Fully automated operation, automatic zeroing through the internal zero trap and optional automatic calibration from an internal H₂O₂ permeation source. Time resolution about 90 s (T90).

Technical Specifications

All specifications at standard conditions (23 °C, 1013 hPa, dry gas).

Measurement
AL2021SC — Hydrogen Peroxide (H₂O₂)
Detection PrincipleEnzymatic UV fluorimetry (peroxidase / p-hydroxyphenylacetic acid), continuous
Measurement Range0.1 ppb – 3000 ppb (gas) · 100 ng/l – 3 mg/l (liquid)
Response Time (T90)~90 s (T90), delay ~300 s
Accuracy2% full scale
Sample Flow1 L/min air · ~0.5 mL/min liquid
Channels / Chambers1 channel · multiple gas sources can be connected via valve cluster
Instrument
General
Dimensions (W × D × H)45 × 15 × 56 cm (19″)
Weight15 kg
Power Supply110 / 230 VAC, 50/60 Hz · 24 VDC on request
Power Consumption50 W
Digital InterfaceRS232 · TCP/IP (RS232 ↔ USB adapter included)