Home › Blog › Background CO in the Tropical Marine Boundary Layer: The AL5001 at Cape Verde

Background CO in the Tropical Marine Boundary Layer: The AL5001 at Cape Verde

Since October 2006 the Cape Verde Atmospheric Observatory has measured carbon monoxide in tropical marine background air with an Aero-Laser AL5001 — continuously, as a WMO-GAW global station, with no end date in sight.

AT
Aero-Laser Team
Aero-Laser GmbH
September 2025 · 1 min read

On the island of São Vicente, at Calhau on the Cape Verde archipelago, the trade winds deliver something rare: air that has crossed thousands of kilometres of open Atlantic. The Cape Verde Atmospheric Observatory (CVAO), operated under the UK’s National Centre for Atmospheric Science (NCAS), uses it to answer a fundamental question — how well does the tropical atmosphere clean itself?

The task

The project monitors the background concentration of CO, alongside other trace gases, in the tropical marine boundary layer to better understand the oxidation capacity of this region. CO concentration there is mainly controlled by the hydroxyl radical (OH) — deviations point to long-range transport events or biomass-burning plumes.

As a WMO-GAW global station, the CVAO has to meet strict criteria:

  • Regional representativeness, free from local pollution influences
  • Infrastructure for long-term observations with greater than 90 % data capture
  • Measurements traceably linked to the GAW CO Primary Standard

The setup

An Aero-Laser AL5001 CO monitor (predecessor of the AL5005) has been running at the site since October 2006. The VUV resonance fluorescence method delivers sub-ppb sensitivity with a response time fast enough to resolve individual plume events — and without the water-vapour cross-sensitivity that complicates IR-based measurements in humid marine air. The University of York additionally operates an AL5002 for shorter field deployments. CO data is submitted in near-real-time to the European MACC atmospheric monitoring project.

No end date

Nearly two decades in, there are no plans for the measurements to stop. That is the quiet benchmark for an instrument in network service: not the spec sheet, but the years of data capture above 90 %. If you are planning a background station of your own, our engineers are happy to discuss integration and calibration strategy — see the AL5005, the current generation of our CO analyzer line.